Milliseconds Photon-to-Photon Latency Projection System for Optogenetics
In recent years, research aiming to elucidate biological phenomena using optogenetics has been actively conducted. Optogenetics aims to understand biological phenomena by applying optical stimulation to cells that have been genetically modified to express photosensitivity, thereby controlling cellular states and biological behaviors. Therefore, our laboratory focuses on Caenorhabditis elegans (C. elegans), whose entire neural network has been completely mapped, making it an excellent research model for clarifying the relationship between neural circuits and behaviors. The system aims to control cellular states and biological behaviors by applying targeted optical stimulation to specific cells of organisms moving freely within a microscope's field of view.
When illuminating moving targets, if the photon-to-photon latency between target recognition and the projection update is high, the target will have already moved by the time the light is projected, causing the stimulation to miss its mark. Conversely, by minimizing this photon-to-photon latency, the target can be illuminated while its movement is still negligible, ensuring the light accurately hits the intended position. Therefore, achieving low photon-to-photon latency is an indispensable feature of this system.
To achieve this, our laboratory has developed a high-speed micro-projection system featuring low photon-to-photon latency using a high-speed DMD (Digital Micromirror Device) to realize optical projection onto dynamic targets. The proposed system consists of two microscopes: a projection microscope connected to an optical system that controls the projected image, and a camera microscope connected to a high-speed camera capable of rapid image acquisition(Figure 1, Figure 2).
In the proposed system, the target is first detected at high speed from the image captured by the camera microscope. After that, a projection image matched to the detected target's position is generated through high-speed image processing. Then, by projecting the generated image through the projection microscope, optical projection accurately targeting the object's position is achieved. A high-speed DMD is used to control the projected image, enabling high-speed projection with low photon-to-photon latency. By changing the mirror angles to control the direction of light reflection, the light rays entering the projection microscope are selected, enabling projection onto any intended position.
To verify the effectiveness of the proposed system, we conducted dynamic projection experiments on a randomly moving object. As a result (see video), the positional relationship between the printed letter "3" and the projected letter "3" is almost always constant. This indicates that the photon-to-photon latency of the system is quite short and the projected image is successfully modified according to the captured image. Furthermore, we demonstrate high-speed dynamic projection experiments targeting the head of C. elegans. The results (see video) verify that the system can track the C. elegans's positional and postural changes with low photon-to-photon latency, enabling accurate and dynamic projection onto the intended target.
Movie: Results of high-speed dynamic projection

Fig1. Appearance and configuration of the proposed system
Fig2. Schematic diagram of the proposed system
References
-
Hiromasa Oku, Atsushi Eda, Toya Fujita: Milliseconds photon-to-photon latency projection system for adaptive optogenetics applications ,The 2024 IEEE International Conference on Robotics and Biomimetics (IEEE ROBIO 2024) in( Bangkok, Thailand. 11/12/2024)/Oral Session,ID 221
-
Atsushi Eda, Toya Fujita, and Hiromasa Oku, "Milliseconds photon-to-photon latency projection system for adaptive optogenetics applications," Opt. Express 33, 20157-20168 (2025)